Everfine
Everfine
Everfine
Home Products EMS Electromagnetic Immunity Test System EMS61000-6B Conducted Immunity Test System
EMS61000-6B Conducted Immunity Test System
Conducted immunity and BCI tests per IEC 61000‑4‑6, GB/T 17626.6, YY 9706, ISO 11452‑4, GB/T 33014.4
Typical application scenarios
EUT size is far smaller than low‑frequency wavelengths, but cables act as antennas susceptible to RF fields, affecting performance. EMS61000‑6B is a conducted immunity test system for RF immunity of electronics in automotive, medical, lighting, power, and more.
Everfine
Main characteristics
1.Integrated test host design – compact structure and easy operation.
2.CDN injection – suitable for power lines and common cables.
3.Electromagnetic clamp injection – an alternative when CDN is not applicable; the EM clamp is a high‑efficiency broadband clamp‑type injection device commonly used for testing multiple unshielded cables.
4.Current clamp injection – suitable for patient coupling cables of medical equipment.
Technical Specifications
ModelEMS61000-6B
Frequency Range150kHz~230MHz
Test Level30Vrms(CDN injection)、10Vrms(Clamp injection)
RF Output PowerMAX 100W
RF Measured Power-45dBm~+35dBm
Optional Current Injection Clamp: Compatible with 100 kHz to 400 MHz BCI (Bulk Current Injection) test
EverfineEverfine

QR code

Everfine

Follow us

EverfineEverfine

Online
consultation

Everfine

WeChat Scan Code Contact

EverfineEverfine

Return
top